Operating procedure:
The lamp operating current should be in the range of ⅓ - ⅔ of the maximum current indicated on the lamp.
After use, the lamp window should be wiped with an alcohol-soaked cotton cloth.
After using the lamp below the melting point of the element, do not touch the lamp until it cools down, so as not to damage the cathode element.
To increase the service life of the lamp, it is necessary to warm up the lamp for 30 minutes before use.
Spectral lamps with polycathode for atomic absorption LT-2 are used in optics, photophysics, spectroscopy, chemical engineering production, medicine and are gas-discharge sources with a linear spectrum of emitters, operating on the principle of arc discharge.
The spectral glow discharge lamp with a hollow cathode LT-2 is intended for use as a radiation source for the spectra of various chemical elements in atomic resonance spectrophotometry equipment operating in continuous and pulsed modes.
Ignition voltage of various elements: all not more than 450V
Power supply voltage with current ripple coefficient not more than 1%, V: 500
Duration of current pulses, µs: 20-1700
Duty cycle of current pulses, rel.units: 1-10