Application:
The portable dental suction system is suitable for assisting saliva aspiration during dental treatment.The portable dental suction system consists of air pump, filter, air - liquid separation device and control switch.
Function:
1, Mainly used in dental hospitals, dental clinics, dental comprehensive treatment machine work, due to the handpiece high-speed rotation generated by the aerosol mixture of suction, can be aerosol, saliva, blood, blood clots and small particles at the same time, to solve the dental chair comes with the suction force is not stable enough, to prevent backflow and other problems, to lift the patient swallowing uncomfortable trouble, to prevent oral bacteria cross-infection, improve the doctor's work efficiency;
2, Instead of Inadequate and unstable suction force of the dental chair
Technical parameters:
1. Model: FM-370X
2. Input power: 220V / 50HZ 110V / 60HZ
3. Power: 280W
4. Current: 2.0A
5. Output power:0.25KW
6.Noise: less than or equal to 50dB
7. Maximum vacuum: -110Mbar
8. Ambient temperature: 5℃-40℃
9. Gross weight: 23kg
10. Package size: 48x44x84cm
Operating environment requirements:
Ambient temperature: 10 ℃, 40 ℃
Relative humidity: less than or equal to 80%
Atmospheric pressure: 500hpa-1060hpa;
Maintenance:
1.Regular maintenance and maintenance of equipment can effectively reduce the occurrence of failure and keep your equipment in good condition
2.Good working condition, extend service life, when cleaning and maintenance, operators must wear masks and waterproof gloves to operate.
3.After each use, in order to avoid bad smell and reduce the possibility of infection, a glass of water should be inhaled after the completion of treatment.
4.In order to avoid bad smell and reduce the possibility of infection after a day's use, non-foaming detergent and disinfectant should be used to clean the equipment pipeline.
5.Clean the particles or debris in the filter at least twice a month to avoid the reduction of suction.
Cleaning
Please keep this product clean. Before cleaning this product, please disconnect the power supply, rub the outer surface of the product with neutral detergent and then clean it with damp soft cloth.
Caution: do not use liquid or inflammable detergent warning: do not damage the power cord when moving this product.